FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3882253 · Received June 18, 2014

Report

Report Number
2955842-2014-03752
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER REPORTED COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE INSTRUMENT PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY WEAR. OTHER CABLES AT WRIST WERE NOT DAMAGED. 48- AN ADDITIONAL FINDING WAS VARIOUS SCRATCHES ON THE DISTAL END OF THE MAIN TUBE SHOWING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER BROKEN PITCH CABLE AND MAIN TUBE SCRATCHES WERE TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS OBSERVED DURING CENTRAL PROCESSING THAT THE PROGRASP FORCEPS INSTRUMENT HAD FRAYED WIRES. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
358534 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M10140108 539

Patients

Seq Age Sex Outcome Treatment
1