ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-17504
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Report Date
- June 16, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/16/2014 WITH THE FOLLOWING FINDINGS:ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. INVESTIGATION FOUND THE BATTERY COMPARTMENT WAS CRACKED BELOW THE GRIP PAD AND THE THREADS AREA. THE VERIFY SCREEN WAS DIM AND DISCOLORED. THE CONTRAST BUTTON WAS UNRESPONSIVE AND THE BUTTON CONTACT WAS FOUND MISSING. IN ADDITION, CONTAMINATION WAS FOUND UNDER THE ¿OK¿, ¿UP¿ AND ¿DOWN¿ BUTTON CONTACTS. UNRELATED TO THE CASING, DISPLAY AND BUTTON ISSUES, IT WAS OBSERVED THAT THE KEYPAD WAS PEELING AND TORN. NO EVIDENCE OF MOISTURE INTRUSION OR LOOSE COMPONENTS WAS FOUND INSIDE THE PUMP. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED, THE DISPLAY WAS DIM/DISCOLORED AND THE BUTTONS HAD MALFUNCTIONED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/16/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358522 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |