FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 1000 DEFIBRILLATOR
MDR report key: 3882189
·
Received June 18, 2014
Report
- Report Number
- 3015876-2014-00674
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K042404
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A BROKEN INTERNAL STRAP FROM THE CAPACITOR FOIL WITHIN THE MAIN HIGH ENERGY CAPACITOR. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WAS SHOWING ITS SERVICE WRENCH AND NEEDED ATTENTION. UPON EVALUATION BY PHYSIO-CONTROL, IT WAS OBSERVED THAT THE DEVICE WOULD NOT DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358658 | LIFEPAK(R) 1000 DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |