FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3882069 · Received June 18, 2014

Report

Report Number
3004753838-2014-04789
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
April 16, 2014
Report Date
May 23, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED, CONFIRMING THE REPORTED EVENT OF NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER. A CAPA HAS BEEN INITIATED FOR THIS ISSUE.

Description of Event or Problem · 1

AN INTERNATIONAL DISTRIBUTOR CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM INTERMITTENT AUDIO OUTPUT ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED.

Description of Event or Problem · 1

AN INTERNATIONAL DISTRIBUTOR CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO AUDIO OUTPUT ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359270 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649-1 5122252

Patients

Seq Age Sex Outcome Treatment
1 Other