FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 3882024 · Received June 18, 2014

Report

Report Number
2031642-2014-00510
Event Type
Malfunction
Date Received
June 18, 2014
Report Date
May 21, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR SHUT DOWN DURING EST. THE DEVICE WAS NOT IN USE ON A PATIENT THEREFORE THERE WAS NO PATIENT INVOLVEMENT OR HARM. THE MANUFACTURERS FIELD SERVICE ENGINEER (FSE) WAS ABLE TO DUPLICATE THE REPORTED PROBLEM AND FOUND THE AC POWER CORD TO FUNCTION INTERMITTENTLY. THE FSE REPLACED THE AC POWER CORD TO CORRECT THE REPORTED PROBLEM. APPLICABLE FINAL TESTING WAS COMPLETED PER OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359291 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1