PDS PLUS ANTIBACTERIAL (POLYDIOXANONE) SUTURE
Report
- Report Number
- 2210968-2014-07844
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 28, 2014
- Report Date
- May 29, 2014
- Manufacturer
- ETHICON INC.
- Product Code
- NEW
- PMA / PMN Number
- K061037
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCLUSION: TO DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED FOR EVALUATION, ANY FURTHER INFO DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
CONCLUSION: THE ACTUAL NEEDLE REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE.
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGICAL PROCEDURE ON (B)(6) 2014 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE. THE NEEDLE PIECE WAS RETRIEVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358898 | PDS PLUS ANTIBACTERIAL (POLYDIOXANONE) SUTURE | SUTURE, SURGICAL, ABSORBABLE | NEW | ETHICON INC. | UNK | UNK |