FDA Adverse Event Malfunction Summary report: N

3 MOTOR ADVANCE BED

MDR report key: 3881781 · Received May 27, 2014

Report

Report Number
1824206-2014-01632
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CABLE WAS JAMMED IN BRAKE BLOCK. PER THE HILL-ROM SERVICE MANUAL THE ADVANCE SERIES BED REQUIRES EFFECTIVE MAINTENANCE PROGRAM. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. TEST THE BRAKE CASTERS TO DETERMINE IF THE BED MOVES WHEN YOU ACTIVATE THE BRAKE MODE, REPLACE OR ADJUST WHEN NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006-2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE BRAKE BEARINGS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN THE BED SHOP AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310905 3 MOTOR ADVANCE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1130

Patients

Seq Age Sex Outcome Treatment