3 MOTOR ADVANCE BED
Report
- Report Number
- 1824206-2014-01632
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Date of Event
- April 30, 2014
- Report Date
- April 30, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE TECHNICIAN FOUND THE BRAKE CABLE WAS JAMMED IN BRAKE BLOCK. PER THE HILL-ROM SERVICE MANUAL THE ADVANCE SERIES BED REQUIRES EFFECTIVE MAINTENANCE PROGRAM. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. TEST THE BRAKE CASTERS TO DETERMINE IF THE BED MOVES WHEN YOU ACTIVATE THE BRAKE MODE, REPLACE OR ADJUST WHEN NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006-2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE BRAKE BEARINGS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN THE BED SHOP AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310905 | 3 MOTOR ADVANCE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1130 |