FDA Adverse Event Malfunction Summary report: N

PRIME CARE (R) TRANSCEND ADVANCED MATTRESS

MDR report key: 3881769 · Received May 27, 2014

Report

Report Number
3007538326-2014-00046
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 14, 2014
Report Date
May 27, 2014
Manufacturer
PRIMUS MEDICAL LLC
Product Code
IKY
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON INSPECTION OF THIS MATTRESS, THE URETHANE COVER BUBBLED AT THE CTR AND FOOT END OF THE MATTRESS COVER BUT DID PEEL AWAY EXPOSING THE FOAM OF THE MATTRESS. THIS MATTRESS HAS BEEN ISOLATED IN THE NON-CONFORMING ROOM UNTIL INVESTIGATION IS COMPLETE AND MATTRESS HAS BEEN DISPOSITIONED. A NEW MATTRESS WAS DELIVERED TO THE FACILITY ON (B)(4)2014. THIS PROBLEM HAS BEEN ASSIGNED TO CAPA (B)(4), AND A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE CORRECTIVE ACTION.

Description of Event or Problem · 1

CUSTOMER CALLED AND SAID THAT THEY HAVE A MATTRESS THAT IS DELAMINATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311536 PRIME CARE (R) TRANSCEND ADVANCED MATTRESS IKY PRIMUS MEDICAL LLC TACM3580 061312

Patients

Seq Age Sex Outcome Treatment
1