HOMECHOICE
Report
- Report Number
- 1416980-2014-19395
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 3, 2014
- Report Date
- May 24, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS EVALUATED AND THE ISSUE WAS IDENTIFIED THROUGH REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCING TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 21:35:50. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 951ML, INDICATING THE HOME PATIENT (HP) DRAINED 951ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 357045 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |