FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3881701 · Received June 18, 2014

Report

Report Number
1416980-2014-19396
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 24, 2014
Report Date
May 24, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USER TO CONNECT THE PATIENT LINE EXTENSION TO THE PATIENT LINE PRIOR TO PRIMING. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A CHECK PATIENT LINE ALARM ON THE HOMECHOICE (HC) MACHINE, IT WAS REPORTED THAT THE HOME PATIENT HAD CONNECTED THE PATIENT LINE EXTENSION AFTER THE PRIMING PHASE WAS COMPLETED ON THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE PATIENT IN ENDING THERAPY AND EXPLAINED THE PROPER SETUP OF THE PATIENT LINE EXTENSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
357654 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 54 YR HOMECHOICE