FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 3881594 · Received February 19, 2014

Report

Report Number
1824206-2014-00511
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
January 22, 2014
Report Date
January 22, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE RUBBER ON THE BRAKE CASTER WHEELS IS WORN CAUSING THE BRAKES TO NOT HOLD WHEN ENGAGED. THE TECH REPLACED THE BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECH REPORTED THAT THE BRAKE CASTERS WILL NOT STOP THE BED FROM ROLLING WHEN THE BRAKE IS APPLIED. THE BED WAS LOCATED IN THE HILL-ROM SVC CTR. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104287 TOTALCARE BARIATRIC BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1