FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 3881592 · Received February 19, 2014

Report

Report Number
1824206-2014-00510
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
January 22, 2014
Report Date
January 22, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE HEAD END BRAKE CASTERS ARE WORN, THE TECH REPLACED THE HEAD END BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECH REPORTED THE BRAKES AT THE HEAD END OF THE BED ARE NOT HOLDING. THE BED IS LOCATED IN THE HILL-ROM SVC CTR. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104557 TOTALCARE BARIATRIC BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1