FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3881591 · Received February 19, 2014

Report

Report Number
1824206-2014-00509
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
January 22, 2014
Report Date
January 22, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE SIDE RAIL LATCHING MECHANISM IS DIRTY AND STICKY FROM FLUID EGRESS. THE TECH CLEANED AND LUBRICATED THE SIDE RAIL LATCH MECHANISM TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECH REPORTED THE SIDE RAIL WOULD NOT LATCH IN THE UP POSITION. THE BED IS LOCATED IN THE BED SHOP AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104234 TOTALCARE BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1