FDA Adverse Event
Injury
Summary report: N
TALENT AAA
MDR report key: 3881518
·
Received June 18, 2014
Report
- Report Number
- 2953200-2014-01214
- Event Type
- Injury
- Date Received
- June 18, 2014
- Date of Event
- May 27, 2014
- Report Date
- May 27, 2014
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT WERE NOT REPORTED. IT WAS REPORTED THAT THE PATIENT WAS IN FOR A ROUTINE FOLLOW UP. A CT NOTED THAT THE DEVICE HAS MIGRATED RESULTING IN A TYPE I ENDOLEAK. THE ANEURYSM HAS GONE FROM 5 CM TO 5.5 CM IN DIAMETER. THE DEVICE MIGRATED 4.5 CM. THE CAUSE OF THE MIGRATION IS DUE TO NECK REMODELING. THE PHYSICIAN TREATED THE PATIENT WITH TWO ENDURANT STENT GRAFTS. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 357447 | TALENT AAA | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V00381462 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00086 YR | Required Intervention |