TALENT AAA

MDR report key: 3881518 · Received June 18, 2014

Report

Report Number
2953200-2014-01214
Event Type
Injury
Date Received
June 18, 2014
Date of Event
May 27, 2014
Report Date
May 27, 2014
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT WERE NOT REPORTED. IT WAS REPORTED THAT THE PATIENT WAS IN FOR A ROUTINE FOLLOW UP. A CT NOTED THAT THE DEVICE HAS MIGRATED RESULTING IN A TYPE I ENDOLEAK. THE ANEURYSM HAS GONE FROM 5 CM TO 5.5 CM IN DIAMETER. THE DEVICE MIGRATED 4.5 CM. THE CAUSE OF THE MIGRATION IS DUE TO NECK REMODELING. THE PHYSICIAN TREATED THE PATIENT WITH TWO ENDURANT STENT GRAFTS. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
357447 TALENT AAA SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V00381462

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Required Intervention