FDA Adverse Event
Malfunction
Summary report: N
ADVANTA BED
MDR report key: 3881498
·
Received February 19, 2014
Report
- Report Number
- 1824206-2014-00521
- Event Type
- Malfunction
- Date Received
- February 19, 2014
- Date of Event
- January 22, 2014
- Report Date
- January 22, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE LEFT HEAD SIDERAIL WAS BENT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006-2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REMOVED ALL THE PARTS FROM THE LEFT HEAD SIDERAIL AND INSTALLED A NEW RAIL FRAME TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE LEFT HEAD SIDERAIL WOULD NOT GO UP OR LATCH. THE BED IS LOCATED IN THE BED SHOP. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104218 | ADVANTA BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1604 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |