FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 3881498 · Received February 19, 2014

Report

Report Number
1824206-2014-00521
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
January 22, 2014
Report Date
January 22, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE LEFT HEAD SIDERAIL WAS BENT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006-2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REMOVED ALL THE PARTS FROM THE LEFT HEAD SIDERAIL AND INSTALLED A NEW RAIL FRAME TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE LEFT HEAD SIDERAIL WOULD NOT GO UP OR LATCH. THE BED IS LOCATED IN THE BED SHOP. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104218 ADVANTA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1