FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3881279 · Received May 27, 2014

Report

Report Number
3004464228-2014-00718
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 20, 2014
Report Date
April 29, 2014
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K122953
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE REPORTED KINKED CANNULA OR TO DETERMINE WHETHER IT CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD'S USER GUIDE WARNS TO "TEST RESULTS GREATER THAN 250 MG/ML MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS OVER 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER."

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT POD WAS DEACTIVATED AND SHE NOTICED THE CANNULA WAS KINKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310555 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 14000 L40814

Patients

Seq Age Sex Outcome Treatment
1 13 YR