FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 3881220
·
Received January 13, 2014
Report
- Report Number
- 2938836-2014-02154
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- August 3, 2012
- Manufacturer
- ST. JUDE MEDICAL INC., CARDIAC RHYTHM MANAGEMENT D
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. THE REPORTED FIELD EVENT OF HELIX ANOMALY AS WELL AS SENSING AND THRESHOLD ANOMALIES WAS NOT CONFIRMED IN THE LABORATORY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT ABOUT TEN DIFFERENT PLACING MANEUVERS WERE UNDERTAKEN BUT SENSING OR THRESHOLD RESULTS WERE NOT ACCEPTABLE. IT WAS SEEN IN FLUOROSCOPY THAT THE SCREW WAS NO LONGER WORKING. THE LEAD WAS NOT IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 24412 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL INC., CARDIAC RHYTHM MANAGEMENT D | 7122Q/65 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |