FDA Adverse Event
Malfunction
Summary report: N
STERLING?
MDR report key: 3881138
·
Received June 18, 2014
Report
- Report Number
- 2134265-2014-03441
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- DQY
- PMA / PMN Number
- K053116
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATED BY MFR: THE DEVICE WAS NOT RETURNED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND MODERATELY TORTUOUS SUPERFICIAL FEMORAL ARTERY. THE 6.0X40MM STERLING BALLOON CATHETER WAS ADVANCED AND DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 10ATMS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT¿S STATUS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358699 | STERLING? | CATHETER, PERCUTANEOUS | DQY | BOSTON SCIENTIFIC - MAPLE GROVE | H74939032604080 | 16604887 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |