FDA Adverse Event Malfunction Summary report: N

CUSA EXCEL 23KHZ STANDARD TIP

MDR report key: 3881096 · Received April 7, 2014

Report

Report Number
3006697299-2014-00019
Event Type
Malfunction
Date Received
April 7, 2014
Date of Event
March 20, 2014
Report Date
March 20, 2014
Manufacturer
INTEGRA LIFESCIENCES IRELAND LIMITED
Product Code
LFL
PMA / PMN Number
K981262
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO.

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED TWO CUSA EXCEL TIPS (PRODUCT ID: C4601, LOT NUMBERS 113630 AND 1133588) WERE USED ON A LIVER TUMOR PROCEDURE AND BOTH TIP ENDS BROKE OFF DURING USE. THE TIP END WAS LOST IN THE SURGICAL SITE FOR A SHORT TIME UNTIL THE SURGEON EVENTUALLY LOCATED IT. THE TIPS WERE NOT IN CONTACT WITH ANY OTHER INSTRUMENT, HOWEVER, THE CEM NOSECONE WAS BEING USED. THERE WAS NO PT INJURY REPORTED AND THE PRODUCT PROBLEM LED TO A 30 MIN INCREASE IN SURGERY TIME. ADD'L INFO WAS THEN RECEIVED ON (B)(6) 2014 WITH THE FOLLOWING: IT WAS CLARIFIED THAT ONLY ONE OF THE TWO CUSA EXCEL TIPS USED HAD ACTUALLY BROKEN OFF. THE USER HAD OPENED ANOTHER TIP TO REPLACE THE BROKEN TIP. THE REPLACEMENT TIP WORKED PERFECTLY WELL. THE USER COULD NOT REMEMBER WHICH PACKAGING AND LOT NUMBER WENT WITH WHICH TIP, HENCE TWO LOT NUMBERS WERE PROVIDED TO THE SALE REP WHEN IT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
208390 CUSA EXCEL 23KHZ STANDARD TIP ULTRASONIC SURGICAL PRODUCTS LFL INTEGRA LIFESCIENCES IRELAND LIMITED 1131630 OR 1133588

Patients

Seq Age Sex Outcome Treatment
1