3.5MM TI LOCKING SCREW SELF-TAPPING 28MM
Report
- Report Number
- 9612488-2014-10232
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 22, 2014
- Manufacturer
- SYNTHES BETTLACH
- Product Code
- KTT
- PMA / PMN Number
- PK000684
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ADDITIONAL NARRATIVE: A MANUFACTURING EVALUATION WAS COMPLETED: THE MICROSCOPIC INVESTIGATIONS OF THE COMPLAINED LOCKING SCREW MANUFACTURED IN OCTOBER 2012 SHOWS THAT THE THREAD ON THE SCREW SHAFT WAS DAMAGED DURING USE. THE TIP OF THE FLANKS WAS PARTIALLY SHEARED OFF RESULTING IN THE VISIBLE BURR. BECAUSE OF THE DAMAGE, THE COMPLAINT RELEVANT DIMENSIONS CANNOT BE CHECKED FOR DIMENSIONAL ACCURACY OF THE VALID MANUFACTURING SPECIFICATIONS. THE ANODIZED GREEN COLOR WAS ABRADED ALMOST ON THE ENTIRE LENGTH. THE FLANKS ARE ROUNDED AND WORN. THE REVIEW OF THE MANUFACTURING DOCUMENTS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE CONDITION OF THE RETURNED DEVICE INDICATES THAT THE SCREW CAME IN CONTACT WITH THE PLATE DURING INSERTION OR EXTRACTION WHILE USE. NO PRODUCT RELATED CONDITION WAS DETECTED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE WAS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. NO NON-CONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT WHILE INSERTING THE 3.5 LOCKING SCREWS TO PHILOS, THE PAINT OF THE SCREW WAS STRIPPED AND LEFT A COIL. THE SURGEON DID IRRIGATION FOR AN ADDITIONAL THIRTY MINUTES. NO PATIENT HARM WAS REPORTED. THIS IS REPORT 2 OF 2 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 357153 | 3.5MM TI LOCKING SCREW SELF-TAPPING 28MM | APPLIANCE, FIXATION, NAIL | KTT | SYNTHES BETTLACH | 8123116 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |