FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 3880821 · Received April 9, 2014

Report

Report Number
2242352-2014-00414
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
March 13, 2014
Report Date
March 17, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, ANCHOR TAB AND THE SEAL WERE NOT RETURNED. THE BLUE SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVAL RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAIL TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION ((B)(4)) ISSUED FOR THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, THE HS III PROXIMAL SEAL FAILED TO DEPLOY. THE SURGEON TRIED TO PULL THE SEAL OUT OF THE OUT DELIVERY DEVICE, BUT THE STRING STAYED INSIDE THE LOADER. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214076 HS III PROXIMAL SEAL SYSTEM 3.8MM ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25091352

Patients

Seq Age Sex Outcome Treatment
1 64 YR