FDA Adverse Event Malfunction Summary report: N

TX1 TISSUE REMOVAL SYSTEM - MICROTIP

MDR report key: 3880742 · Received April 9, 2014

Report

Report Number
2085033-2014-00557
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
January 15, 2014
Report Date
April 9, 2014
Manufacturer
TENEX HEALTH
Product Code
LFL
PMA / PMN Number
K123640
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE USER OBSERVED A CRACK IN THE CLEAR CASING. EVALUATION OF THE RETURNED UNIT CONFIRMED A CRACK IN THE PLASTIC CASING COMPROMISING UNIT STERILITY. NO STRESS MARKS INDICATING INTERNAL PRODUCT CONTENTS CAUSED CRACK OR PUNCTURE WERE OBSERVED. THE UNIT WAS RETURNED IN THE ORIGINAL TRAY WITH THE SEAL STILL ATTACHED. THE UNIT WAS NEVER USED ON A PATIENT. THIS COMPLAINT IS STILL UNDER INVESTIGATION TO DETERMINE THE ROOT CAUSE OF THE CRACK. CURRENTLY BELIEVED TO BE CAUSED BY ENVIRONMENTAL CONDITIONS (I.E. TEMPERATURE IN EXTREME COLD).

Description of Event or Problem · 1

THE NURSE OPENED THE BOX AND PULLED OUT THE TX KIT AND THE PLASTIC WAS BROKEN MAKING THE DEVICE NON STERILE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213350 TX1 TISSUE REMOVAL SYSTEM - MICROTIP ULTRASONIC SURGICAL ASPIRATOR LFL TENEX HEALTH 32613-01

Patients

Seq Age Sex Outcome Treatment
1