FDA Adverse Event Malfunction Summary report: N

HX2 TEMPERATURE MANAGEMENT SYSTEM

MDR report key: 3880705 · Received April 9, 2014

Report

Report Number
1828100-2014-00251
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
March 19, 2014
Report Date
March 19, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K071521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) RESEATED THE HEATER CONNECTORS AND RETESTED THE UNIT. LEAKAGE CURRENT IS WITHIN SPECIFICATIONS AND ALL FUNCTIONS PASSED. THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE HEATER COOLER UNIT EXPERIENCED AN EXCESSIVE LEAKAGE CURRENT. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213339 HX2 TEMPERATURE MANAGEMENT SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 809810

Patients

Seq Age Sex Outcome Treatment
1