HX2 TEMPERATURE MANAGEMENT SYSTEM
MDR report key: 3880705
·
Received April 9, 2014
Report
- Report Number
- 1828100-2014-00251
- Event Type
- Malfunction
- Date Received
- April 9, 2014
- Date of Event
- March 19, 2014
- Report Date
- March 19, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K071521
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) RESEATED THE HEATER CONNECTORS AND RETESTED THE UNIT. LEAKAGE CURRENT IS WITHIN SPECIFICATIONS AND ALL FUNCTIONS PASSED. THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
Description of Event or Problem · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE HEATER COOLER UNIT EXPERIENCED AN EXCESSIVE LEAKAGE CURRENT. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.