FDA Adverse Event
Malfunction
Summary report: N
VERSACARE P500 NSC AIR RENTAL BED
MDR report key: 3880499
·
Received May 30, 2014
Report
- Report Number
- 1824206-2014-01673
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- May 3, 2014
- Report Date
- May 3, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH FOUND THAT THE SCALE WAS ZEROED WITH A PT IN THE BED CAUSING BOTH ISSUES, USER ERROR. THE TECH ZEROED THE SCALE WITHOUT A PT IN THE BED AND IN-SERVICED THE NURSE ON THE BED EXIT FUNCTIONS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
THE ACCOUNT REPORTED THE BED EXIT WOULD NOT SET AND THE SCALE WAS NOT FUNCTIONING. THE BED WAS LOCATED IN ROOM 613 AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319884 | VERSACARE P500 NSC AIR RENTAL BED | A/C POWERED ADJUSTABLE HOSP BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |