FDA Adverse Event Malfunction Summary report: N

VERSACARE P500 NSC AIR RENTAL BED

MDR report key: 3880499 · Received May 30, 2014

Report

Report Number
1824206-2014-01673
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 3, 2014
Report Date
May 3, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THAT THE SCALE WAS ZEROED WITH A PT IN THE BED CAUSING BOTH ISSUES, USER ERROR. THE TECH ZEROED THE SCALE WITHOUT A PT IN THE BED AND IN-SERVICED THE NURSE ON THE BED EXIT FUNCTIONS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BED EXIT WOULD NOT SET AND THE SCALE WAS NOT FUNCTIONING. THE BED WAS LOCATED IN ROOM 613 AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319884 VERSACARE P500 NSC AIR RENTAL BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1