DIMENSION(R) CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2014-00331
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 20, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
- Product Code
- JJE
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE OF BIASED HIGH HB1C RESULTS IS UNKNOWN. THE CAUSE OF THE BIASED HIGH RESULTS IS UNDER INVESTIGATION. THE HEADQUARTERS SUPPORT CENTER HAS DIRECTED THE CUSTOMER SERVICE ENGINEER TO INVESTIGATE SEVERAL INSTRUMENT COMPONENTS. RECOMMENDATIONS ON CALIBRATOR PREPARATION AND HANDLING HAVE BEEN MADE TO OPTIMIZE INSTRUMENT TO INSTRUMENT AGREEMENT.
BIASED HIGH HEMOGLOBIN A1C (HB1C) RESULTS WERE OBTAINED ON QC AND PATIENT SAMPLES ON DIMENSION EXL INSTRUMENT (B)(4). QC RESULTS HOWEVER, REMAINED WITHIN UPPER LIMITS OF THE CUSTOMER LABORATORY RANGES. LOWER QC RESULTS WITHIN LABORATORY RANGES AND LOWER PATIENT RESULTS WERE OBTAINED ON AN ALTERNATE DIMENSION INSTRUMENT AND WITH AN ALTERNATE REFERENCE METHOD. THE PATIENT RESULTS WERE REPORTED TO THE PHYSICIAN DURING THE INTERVAL OF BIASED HIGH RESULTS. THERE IS NO INDICATION OF RESULTS HAVING BEEN QUESTIONED BY PHYSICIANS. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS THE BASED HIGH HB1C RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE BASED HIGH HB1C RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358014 | DIMENSION(R) CLINICAL CHEMISTRY SYSTEM | DIMENSION(R) EXL CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD | DIMENSION EXL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |