FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 3880206 · Received January 13, 2014

Report

Report Number
2938836-2014-02165
Event Type
Injury
Date Received
January 13, 2014
Date of Event
December 18, 2012
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INADEQUATE ATP AND SHOCKS WHILE IN SVT. NOISE WAS OBSERVED ON THE VENTRICULAR LEAD. THE NOISE WAS NOT REPRODUCIBLE AT THE CLINIC. DAMAGE LEADS, CONNECTION ISSUE OR OTHER METAL AGAINST METAL CONTACT WERE SUSPECTED. THE LEAD REMAINED IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27044 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD, NVY NVY ST. JUDE MEDICAL, INC., CRMD 7122Q/65

Patients

Seq Age Sex Outcome Treatment
1 62 YR