FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3880199 · Received June 18, 2014

Report

Report Number
3004209178-2014-85670
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
April 18, 2014
Report Date
May 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UNABLE TO PRIME DURING THE PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO PRIME ALARM NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, BROKEN RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE NEAR RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS AND BROKEN INSULIN PUMP BELT CLIP SLOT. THE INSULIN PUMP HAD AN INK SPOT/BLEEDING ON MIDDLE ON LCD GLASS.

Description of Event or Problem · 1

IT WAS REPORTED BY NURSE THAT CUSTOMER WAS HAVING TROUBLE WITH HIS INSULIN PUMP. CUSTOMER IS AN INMATE AT A FEDERAL FACILITY. THE INSULIN PUMP IS ALARMING. THE BATTERY HAS BEEN CHANGED; THE INSULIN PUMP WILL NOT COME BACK ON. THE CURRENT BLOOD GLUCOSE READING IS 35 MG/DL. CUSTOMER HAS TREATED WITH FOOD. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.