FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3880195
·
Received April 9, 2014
Report
- Report Number
- 2518422-2014-00568
- Event Type
- Malfunction
- Date Received
- April 9, 2014
- Date of Event
- March 19, 2014
- Report Date
- March 19, 2014
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MFR RECEIVED INFO ALLEGING A VENTILATOR WAS ALARMING FOR A LOW INSPIRATORY PRESSURE CONDITION. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214085 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1054655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |