FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3880155 · Received January 30, 2014

Report

Report Number
1314492-2014-05103
Event Type
Malfunction
Date Received
January 30, 2014
Date of Event
January 1, 2014
Report Date
January 3, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION EXPERIENCED THE REPORTED SYSTEM ERROR 322. A PHYSICAL INSPECTION FOUND THAT THE UPPER LATCH SWITCH TO MOVE IN AND OUT FROM THE UPPER DOOR LATCH. THE LOOSE SCREWS WILL TRIGGER AN INTERMITTENT OPEN/CLOSED SWITCH CONNECTION CAUSING THE SYSTEM ERROR 322. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP HAS A SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65412 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1