FDA Adverse Event Injury Summary report: N

FORTIFY DR, DF-4 CONNECTOR

MDR report key: 3880119 · Received January 13, 2014

Report

Report Number
2938836-2014-02247
Event Type
Injury
Date Received
January 13, 2014
Date of Event
November 29, 2011
Manufacturer
ST. JUDE MEDICAL INC., CRMD
Product Code
LWS
PMA / PMN Number
P910023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT RECEIVED INAPPROPRIATE ATP AND SHOCK THERAPY. TECH SERVICE RECOMMENDED DISABLING THE ATRIOVENTRICULAR INTERVAL DELTA AND INCREASING THE VENTRICULAR FIBRILLATION ZONE. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
24300 FORTIFY DR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, LWS LWS ST. JUDE MEDICAL INC., CRMD CD2231-40Q

Patients

Seq Age Sex Outcome Treatment
1 65 YR