FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3880118
·
Received January 30, 2014
Report
- Report Number
- 1314492-2014-05098
- Event Type
- Malfunction
- Date Received
- January 30, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 3, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO.: (B)(4). THE DEVICE WAS NOT RETURNED TO BAXTER FOR EVALUATION AND THEREFORE AN EVALUATION COULD NOT BE COMPLETED. IF THE DEVICE IS RETURNED, AN EVALUATION WILL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE TESTING, A PUMP WAS UNDER INFUSING. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 65408 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |