FDA Adverse Event
Injury
Summary report: N
EPIC PLUS VR
MDR report key: 3880096
·
Received January 13, 2014
Report
- Report Number
- 2938836-2014-02280
- Event Type
- Injury
- Date Received
- January 13, 2014
- Date of Event
- February 20, 2012
- Manufacturer
- ST. JUDE MEDICAL INC., CRMD
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SYMPTOMATIC PATIENT PRESENTED IN CLINIC DURING FOLLOW UP WITH A DEVICE THAT WAS POST SENSED TO WAVE OVERSENSING. THE PATIENT RECEIVED INAPPROPRIATE HIGH VOLTAGE THERAPY AND THE OVERSENSING WAS NOTED ON A STORED ELECTROGRAM. REPROGRAMMING CHANGES WERE RECOMMENDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 24866 | EPIC PLUS VR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, LWS | LWS | ST. JUDE MEDICAL INC., CRMD | V-196 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |