FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3880049
·
Received January 31, 2014
Report
- Report Number
- 1314492-2014-05285
- Event Type
- Malfunction
- Date Received
- January 31, 2014
- Date of Event
- January 6, 2014
- Report Date
- January 6, 2014
- Manufacturer
- BAXTER HLTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER REF: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. EVAL REPRODUCED AND CONFORMED THE REPORTED SYMPTOM OF SYSTEM ERROR 322. PHYSICAL INSPECTION FOUND THAT THE UPPER LATCH SWITCH BRACKET SCREWS WERE LOOSE ALLOWING THE UPPER LATCH SWITCH TO MOVE IN AND OUT FROM THE UPPER DOOR LATCH. THE LOOSE SCREWS WILL TRIGGER AN INTERMITTENT OPEN/CLOSED SWITCH CONNECTION CAUSING THE SYMPTOM. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED. SYS ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP HAD A SYSTEM ERROR 322. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67450 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HLTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |