FDA Adverse Event
Injury
Summary report: N
TRANSCUTANEOUS BLOOD GAS MONITOR
MDR report key: 3879434
·
Received May 29, 2014
Report
- Report Number
- 1523456-2014-00002
- Event Type
- Injury
- Date Received
- May 29, 2014
- Date of Event
- April 1, 2014
- Report Date
- May 2, 2014
- Manufacturer
- RADIOMETER MEDICAL APS
- Product Code
- LKD
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
THE PATIENT (A NEONATAL BOY BORN IN (B)(6) GESTATION WEEK) RECEIVED A BURN MARK ON THE RIGHT THIGH. THE TEMPERATURE WAS SET TO 43 DEGREES CELSIUS, AND THE ELECTRODE WAS PLACED ON THE SAME LOCATION FOR 7 HOURS. ACCORDING TO THE OPERATOR'S MANUAL (TCM4/40 OM PAGE 6-12) THE PROBE SHOULD BE RELOCATED AFTER 4 HOURS AT LEAST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316271 | TRANSCUTANEOUS BLOOD GAS MONITOR | TCM4 MONITORING SYSTEM | LKD | RADIOMETER MEDICAL APS | TCM4 SERIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |