FDA Adverse Event Other Summary report: N

MOTHERMATES HYDROGEL BREAST PAD INTNL

MDR report key: 3878064 · Received June 3, 2014

Report

Report Number
1219103-2014-00013
Event Type
Other
Date Received
June 3, 2014
Date of Event
May 15, 2014
Report Date
May 16, 2014
Manufacturer
COVIDIEN
Product Code
MGQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION IS CURRENTLY UNDERWAY, UPON COMPLETION THE RESULTS WILL BE FORWARDED.

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN ON (B)(6) 2014 THAT A CUSTOMER HAD AN ISSUE WITH A BREAST PAD. THE CUSTOMER STATES THAT DURING THE REMOVAL OF THE FOIL, THE PADS RIP LEAVING SMALL RESIDUAL PIECES ON THE FOIL AND SOMETIMES ON THE PATIENT'S BREAST. AN INFANT HAD SMALL PARTS OF THE PAD IN ITS MOUTH. ALSO, THE SHEET GETS POROUS AFTER ABOUT THREE DAYS OF USE. THE DURATION OF USE WAS TWO DAYS. THERE WERE NO CONSEQUENCES TO THE PATIENT AS A RESULT. NO ALLERGIC REACTION AND THE PIECES WERE REMOVED FROM THE INFANT'S MOUTH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
325281 MOTHERMATES HYDROGEL BREAST PAD INTNL BREAST PAD MGQ COVIDIEN 9403 317806X

Patients

Seq Age Sex Outcome Treatment
1 UNK Other