FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 3876136 · Received May 20, 2014

Report

Report Number
2242352-2014-00583
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 11, 2014
Report Date
April 23, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE AND SLIGHT EVIDENCE OF BLOOD. THE DELIVERY SERVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE SEAL, TENSION SPRING ASSEMBLY AND THE ANCHOR TAB WERE INSIDE THE LOADING DEVICE, NOT IN THE DELIVERY DEVICE. THE SEAL WAS VISIBLE UNDER THE WINDOW OF THE LOADING DEVICE. THE BLUE SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAILURE TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION (WY-2013-35) ISSUED FOR THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING FAILED TO DEPLOY PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298511 HS III PROXIMAL SEAL SYSTEM 3.8MM ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25094393

Patients

Seq Age Sex Outcome Treatment
1 82 YR