FDA Adverse Event Malfunction Summary report: N

VASO VIEW HEMOPRO

MDR report key: 3876121 · Received May 20, 2014

Report

Report Number
2242352-2014-00581
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 7, 2014
Report Date
April 22, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE BUT THERE WAS EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE SILICONE LAYER OF THE BALLOON HAD RUPTURED. EACH BTT UNDERGOES A FULL INFLATION AND DEFLATION INSPECTION PRIOR TO DISTRIBUTION. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION (WY-2013-94) ISSUED FOR THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE CUFF ON THE BTT RUPTURED ON THE VASO VIEW HEMOPRO. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298506 VASO VIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25093402