FDA Adverse Event Injury Summary report: N

PERMOBIL

MDR report key: 3875142 · Received May 8, 2014

Report

Report Number
1221084-2014-00010
Event Type
Injury
Date Received
May 8, 2014
Date of Event
April 22, 2014
Report Date
May 7, 2014
Manufacturer
PERMOBIL, INC.
Product Code
ITI
PMA / PMN Number
K041219
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PERMOBIL REP (B)(4) HAS BEEN TO THE SISTER'S HOME TO SEE THE RAMP. THE RAMP IS 8FT LONG AND IS GOING UP STEPS WHICH RISE 22INCHES. THE ADA SPECIFICATION FOR THIS RAMP WOULD BE 1:12 OR 22 FEET. THE RECOMMENDATIONS I SEE IN THE INDUSTRY FOR AN OCCUPIED CHAIR ON A RESIDENTIAL RAMP IS 2:12 WHICH WOULD BE 11 FEET AND FOR AN UNOCCUPIED CHAIR IS 3:12 OR 8 FEET. THE ANGLE THAT THE RAMP CURRENTLY IS AT, IS 13 DEGREES. THE PERMOBIL MANUAL CLEARLY STATES THAT 10 DEGREES IS THE MAXIMUM SAFE ANGLE. THE CHAIR LEFT PERMOBIL IN COMPLIANCE WITH THE DMR THE CHAIR "TIPPED" BECAUSE OF DRIVER ERROR THE SLOPE OF THE RAMP IS BEYOND THE SPECIFICATION FOR THE CHAIR.

Description of Event or Problem · 1

CLIENT REPORTS THAT HE WAS GOING DOWN A TEMPORARY PORTABLE RAMP AT HIS SISTER'S HOUSE WHEN HIS CHAIR TIPPED FORWARD. HE RECEIVED A BROKEN LEG IN THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
278056 PERMOBIL POWER WHEELCHAIR ITI PERMOBIL, INC. C300

Patients

Seq Age Sex Outcome Treatment
1 Other