CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2014-02760
- Event Type
- Malfunction
- Date Received
- June 16, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 19, 2014
- Manufacturer
- MDT SOFAMOR DANEK PUERTO RICO MFG
- Product Code
- KWP
- PMA / PMN Number
- K981676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFO: MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE THREAD. FUNCTIONAL EVALUATION WITH SAMPLE MAS HEAD FOUND SET SCREW UNABLE TO BE FULLY ENGAGED. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE MAS AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.
(B)(4). NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT THE SET SCREW SLIPPED DURING INSERTION INTO THE BONE SCREW. THE SCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352770 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | MDT SOFAMOR DANEK PUERTO RICO MFG | NA | H13E2968 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00065 YR | BONE SCREW |