FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3874890 · Received June 16, 2014

Report

Report Number
3005075853-2014-04017
Event Type
Malfunction
Date Received
June 16, 2014
Date of Event
April 29, 2014
Report Date
April 30, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED WITH THE CLAMP ARM DETACHED FROM THE INNER TUBE, BUT FIRMLY FIXED TO THE OUTER TUBE AT THE CLAMP ARM WELD. THE TISSUE PAD WAS FOUND IN GOOD PHYSICAL CONDITION AND PROPERLY ATTACHED TO THE CLAMP ARM DUE TO THE RETURNED CONDITION, THE DEVICE COULD NOT BE TESTED WITH A GENERATOR. POSSIBLE CAUSES OF THE CLAMP ARM DETACHMENT ARE NOT CLOSING THE TRIGGER WHEN SLIDING THE TORQUE WRENCH ON AND OFF; NOT CLOSING THE TRIGGER WHEN INTRODUCING OR REMOVING THROUGH THE TROCAR; OR POSSIBLE ENTANGLEMENT IN FIBROUS TISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE SCREW AT THE END OF THE JAW WAS LOOSE. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.