TRIAGE BNP TEST FOR BECKMAN COULTER
Report
- Report Number
- 2027969-2014-00370
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 31, 2014
- Report Date
- March 31, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- NBC
- PMA / PMN Number
- K052789
- Removal / Correction Number
- 2027969-03/20/14/002C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS ISSUE IS BEING ADDRESSED UNDER (B)(4). INVESTIGATION PENDING.
CAPA 14-009 WAS INITIATED DUE TO AN INCREASE IN BIO-RAD AND OTHER THIRD PARTY CONTROL COMPLAINTS OF LOT TO LOT PATIENT SAMPLE BIAS ASSOCIATED WITH TWO AFFECTED BECKMAN TRIAGE BNP FOR BCI REAGENT PACK LOTS. THE ROOT CAUSE WAS DUE TO A NEW GOAT ANTI-BIOTIN ANTIBODY POOL WHICH CAUSED PATIENT SAMPLES AND SPECIFIC COMMERCIALLY AVAILABLE THIRD PARTY CONTROLS TO RECOVER DIFFERENTLY COMPARED TO PREVIOUSLY RELEASED REAGENT LOTS. THE QC TESTING PROCESS AND SAMPLES USED DID NOT DETECT THESE DIFFERENCES BECAUSE THEY PERFORMED DIFFERENTLY LOT TO LOT THAN PATIENT SAMPLES AND SPECIFIC ADDITIONAL COMMERCIAL CONTROLS. ACTIONS INVOLVED DISCONTINUING THE PURIFIED LOT OF GOAT ANTI-BIOTIN USED IN THE AFFECTED LOTS AND QUALIFYING A NEW LOT. IN ADDITION, PATIENT SAMPLE CORRELATION SET TESTING USED FOR THE QUALIFICATION OF THE NEW GOAT ANTI-BIOTIN WILL BE IMPLEMENTED AS PART OF THE RELEASED TESTING OF SUBSEQUENT REAGENT LOTS. COMPLAINT RATES WERE MONITORED FOR OVER 6 MONTHS WITH SIGNIFICANTLY REDUCED RATES, THEREFORE, THE CAPA WAS DEEMED EFFECTIVE AND CLOSED.
CUSTOMER REPORTED THAT BNP CONTROLS ARE RECOVERING LOW ON LOT #329599. AVERAGE RECOVERY IS ABOUT 10% LOWER. AVERAGE FOR LOT #329599: 33 320 1207. AVERAGE FOR LOT #372048: 36 358 1295.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 250465 | TRIAGE BNP TEST FOR BECKMAN COULTER | B-TYPE NATRIURETIC PEPTIDE TEST | NBC | ALERE SAN DIEGO, INC. | 98200 | 329599 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | BIO-RAD CONTROLS, SECOND LOT #372048 |