FDA Adverse Event Injury Summary report: N

PUMP IN STYLE (ITEM #UNK)

MDR report key: 3874804 · Received June 2, 2014

Report

Report Number
1419937-2014-00446
Event Type
Injury
Date Received
June 2, 2014
Date of Event
May 14, 2014
Report Date
May 15, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. ATTEMPTS TO CONTACT THE CUSTOMER ARE INCOMPLETE AND ONGOING. AS OF THE DATE OF THIS REPORT, THE ORIGINAL PRODUCT HAS NOT BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW UP REPORT WILL BE FILED AT THAT TIME. THE ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PRODUCT THE TRANSFORMERS DURING SHIPPING. IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER'S MASTITIS. REPORTED ISSUES OF MASTITIS ARE UNDER INVESTIGATION IN (B)(4). "MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OF MASTITIS." RIORDAN & WAMBACH, 4TH ED. P. 294: BREASTFEEDING AND HUMAN LACTATION. MASTITIS REQUIRES PROMPT MEDICAL ATTENTION FOR THE MOTHER FOR PAIN RELIEF AND PRESCRIPTION ANTIBIOTICS TO AVOID PROGRESSION TO OVERWHELMING SEPSIS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE ADAPTER FOR THEIR PUMP IN STYLE DEVICE POPPED IN HALF, EXPOSING THE INNER ELECTRONICS, WHICH IS A SAFETY ISSUE. THE CUSTOMER ALSO REPORTED THAT SHE HAS MASTITIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321174 PUMP IN STYLE (ITEM #UNK) HGX MEDELA, INC. UNK/9207010 1212 / REV M

Patients

Seq Age Sex Outcome Treatment
1