FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 3874646
·
Received April 24, 2014
Report
- Report Number
- 1720753-2014-03648
- Event Type
- Malfunction
- Date Received
- April 24, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 24, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CPU PCB WAS EVALUATED AND RESEATED. THE SBC COIN BATTERY VOLTAGE AND POWER SUPPLY VOLTAGE WAS ALSO CONFIRMED TO BE WITHIN SPECIFICATION. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 248569 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |