FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3874605
·
Received April 22, 2014
Report
- Report Number
- 1720753-2014-03518
- Event Type
- Malfunction
- Date Received
- April 22, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 22, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE BACKPLANE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THE SYSTEM HAD A CINE DISK NOT AVAILABLE ERROR MESSAGE ON BOOT UP. THIS WOULD PREVENT THE CINE FUNCTION FROM OPERATING, CAUSING A LOSS OF SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR CINE FUNCTIONS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243508 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |