FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER EDITI

MDR report key: 3874498 · Received April 22, 2014

Report

Report Number
9615050-2014-02863
Event Type
Malfunction
Date Received
April 22, 2014
Date of Event
March 25, 2014
Report Date
March 25, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MECHANISM IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DISTAL PRESSURE SENSOR PIN WAS BROKEN OFF. THE DEVICE WAS RETURNED TO BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243061 PLUM A+ DRIVER EDITI 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE: LIST # UNK, SN UNK