GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00507
- Event Type
- Malfunction
- Date Received
- June 16, 2014
- Date of Event
- May 16, 2014
- Report Date
- June 4, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITED SEVERE BURNT ON DETRITUS; THE GLASS CAP EXHIBITED SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE OUTER FLOW TUBING EXHIBITED MILD CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.
(B)(4).
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE THE FIBER WAS NOTED TO HAVE BEEN "FIRING OUT OF END OF FIBER" THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE, AN ADDITIONAL FIBER WAS NOT NECESSARY SINCE THE FAILURE OCCURRED AT THE END OF THE PROCEDURE. PATIENT OUTCOME: PATIENT IS GOOD AND SURGERY WAS COMPLETED, NO INJURIES OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352500 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 406A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |