FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3874401 · Received June 16, 2014

Report

Report Number
2531779-2014-17077
Event Type
Malfunction
Date Received
June 16, 2014
Report Date
June 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/21/2014 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/10/2014 WITH THE FOLLOWING FINDINGS: THE DATA IN THE BLACK BOX FOR THE DATE OF THE COMPLAINT WAS OVERWRITTEN DUE TO CONTINUED USE OF THE PUMP. DURING TESTING, THE PUMP PERFORMED THE REWIND, LOAD, AND PRIME STEPS WITH NO ALARMS OCCURRING. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS OCCURRING. THE FORCE SENSOR CALIBRATION READING WAS FOUND TO BE OUT OF THE REQUIRED SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND THE FORCE SENSOR RESISTANCE READING WAS FOUND TO BE WITHIN THE REQUIRED SPECIFICATIONS. THE COMPLAINT OF THE PUMP EMITTING MULTIPLE LOSS OF PRIME ALARMS WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT THE PUMP WAS EMITTING MULTIPLE LOSS OF PRIME WARNINGS. IT WAS NOTED THAT THE LOSS OF PRIME WARNINGS HAD OCCURRED WITH AT LEAST THREE SEPARATE CARTRIDGES FROM TWO SEPARATE BOXES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353187 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 51 YR