ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-17093
- Event Type
- Malfunction
- Date Received
- June 16, 2014
- Report Date
- June 11, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 05/13/2015-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP BLACK BOX DATA REVEALED THAT INFORMATION FROM THE EVENT DATE HAD BEEN OVERWRITTEN DUE TO CONTINUED USE. DURING A VISUAL INSPECTION OF THE PUMP, THE BATTERY COMPARTMENT WAS OBSERVED TO BE INTACT WITH NO DAMAGE OR EVIDENCE OF MOISTURE INGRESS. THE BATTERY CAP SECURED PROPERLY TO THE PUMP, AND A POWER LOSS WAS NOT OBSERVED. CURRENT DRAWS MEASURED WITHIN SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF MOISTURE INGRESS OR LOOSE COMPONENTS WAS FOUND.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. IT WAS REPORTED THAT THE PUMP BATTERY LIFE DID NOT LAST AS LONG AS EXPECTED. MULTIPLE BATTERIES FROM DIFFERENT PACKAGES WERE USED IN THE PUMP; HOWEVER, THE ISSUE OF SHORT BATTERY LIFE RECURRED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353401 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |