FDA Adverse Event Malfunction Summary report: N

BP ADVANCED PERSONAL DOUBLE

MDR report key: 3874189 · Received April 23, 2014

Report

Report Number
1419937-2014-00382
Event Type
Malfunction
Date Received
April 23, 2014
Date of Event
March 24, 2014
Report Date
April 18, 2014
Manufacturer
MEDELA INC
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER ADAPTER WAS SENT TO THE CUSTOMER AND REQUESTED THE DEFECTIVE POWER ADAPTER BE RETURNED FOR EVALUATION. THE PRODUCT EVALUATION CONFIRMED THE CUSTOMER REPORTED PROBLEM DAMAGED TRANSFORMER PRONG. THE PRODUCT EVALUATION ALSO REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY, WHICH IS A SAFETY RISK. EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED DAMAGED PRONGS AND THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 13.60 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE, AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 54.90 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE POWER ADAPTER PRONG IS DAMAGED. DURING THE PRODUCT EVALUATION ON 04/08/2014, THE TECHNICIAN FOUND A BREACHED TRANSFORMER HOUSING WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245044 BP ADVANCED PERSONAL DOUBLE HGX MEDELA INC 57065/9207010 1413 REV N

Patients

Seq Age Sex Outcome Treatment
1