FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3873812
·
Received January 27, 2014
Report
- Report Number
- 1314492-2014-04662
- Event Type
- Malfunction
- Date Received
- January 27, 2014
- Date of Event
- December 1, 2013
- Report Date
- December 31, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER REF NO.: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE REPORTED SYMPTOM OF SYSTEM ERROR 322 WAS CONFIRMED THROUGH REVIEW OF THE EVENT HISTORY LOG. IT HAS BEEN DETERMINED THAT THE UPPER AND LOWER AUXILIARY WERE THE FAILING COMPONENTS WHICH CAUSED THIS ERROR. THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP HAD A SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58646 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |